Accueil - Chromatographie Liquide -

Analyse de Cations  

Applications:Anions and Cations

Cations(Simultaneous separation of monovalent and divalent cations: YS-50, YK-421)

Cations and Alkylamines (YS-50)


Cation standards and alkylamines can be analyzed simultaneously with IC YS-50.

icchromato.gif
Sample : 10micro-L
1. Li+ 1ppm
2. Na+ 5ppm
3. NH4+ 5ppm
4. Methylamine 5ppm
5. K+ 20ppm
6. Trimethylamine, TMA 10ppm
7. Triethylamine, TEA 10ppm
8. Mg2+ 10ppm
9. Ca2+ 20ppm

calibrationcurve.gifCorrelation Coefficient
1ppm to 100ppm 
Methylamine 1.000
Trimethylamine 0.9998
Triethylamine 0.9997

Column       : Shodex IC YS-50 (4.6mmID*125mm)
Eluent       : 4mM Methanesulfonic acid aq.
Flow rate    : 1.0mL/min
Detector     : Suppressed conductivity
Column temp. : 40°C

 
 

 